Files
tono/docs/nodes/Tip Deconvolution.md

27 lines
877 B
Markdown

# Tip Deconvolution
Reconstruct the true surface from a tip-broadened measured AFM image. Uses morphological grey erosion (Villarrubia algorithm). Equivalent to gwy_tip_erosion (tip.c).
## Inputs
| Name | Type | Required | Description |
|------|------|----------|-------------|
| field | DATA_FIELD | Yes | Measured AFM image |
| tip | DATA_FIELD | Yes | Tip shape field, e.g. from Tip Model or Blind Tip Estimate |
## Outputs
| Name | Type | Description |
|------|------|-------------|
| surface | DATA_FIELD | Reconstructed true surface |
## Controls
None.
## Notes
- The tip pixel size must match the image pixel size exactly; mismatched calibrations will produce incorrect results.
- Deconvolution can only reduce tip broadening; it cannot recover information lost below the noise floor.
- Very large tip sizes are slow due to the nested loop over all tip pixels.