# PFM Analysis
Compute polarization vectors from Piezoresponse Force Microscopy (PFM) data by combining vertical (VPFM) and lateral (LPFM) amplitude and phase channels into polarization magnitude, azimuth, and inclination maps. Equivalent to Gwyddion's `pfm.c` module.
## Inputs
| Name | Type | Required | Description |
|------|------|----------|-------------|
| vpfm_amplitude | DATA_FIELD | Yes | Vertical PFM amplitude channel |
| lpfm_amplitude | DATA_FIELD | Yes | Lateral PFM amplitude channel |
| vpfm_phase | DATA_FIELD | Yes | Vertical PFM phase channel (radians) |
| lpfm_phase | DATA_FIELD | Yes | Lateral PFM phase channel (radians) |
## Outputs
| Name | Type | Description |
|------|------|-------------|
| magnitude | DATA_FIELD | Polarization vector magnitude |
| azimuth | DATA_FIELD | In-plane polarization angle |
| inclination | DATA_FIELD | Out-of-plane polarization angle (3D mode only) |
## Controls
| Name | Type | Default | Description |
|------|------|---------|-------------|
| mode | dropdown | 2d | Decomposition mode: 2d (in-plane only) or 3d (full polarization vector) |
| lateral_sensitivity | FLOAT | 1.0 | Scale factor applied to the lateral signal to compensate for differing VPFM/LPFM detector sensitivities |
## Notes
- PFM measures the local piezoelectric response of ferroelectric materials by detecting surface deformation under an applied AC bias via the AFM cantilever.
- In **2d** mode, only in-plane polarization is resolved from the LPFM channels; magnitude and azimuth are computed but inclination is not produced. In **3d** mode, both vertical and lateral channels are combined to reconstruct the full polarization vector, including inclination.
- The **lateral_sensitivity** parameter compensates for the fact that vertical and lateral deflection detectors typically have different sensitivities. Set this to the ratio of LPFM to VPFM detector sensitivity for accurate vector reconstruction.
- Phase inputs must be in radians. If your data is in degrees, convert before connecting (multiply by pi/180).
- Amplitude channels should be unsigned (absolute) piezoresponse magnitudes; the phase channels encode the sign (direction) of the polarization.